The QUASAR™ IsoCenter Cube is designed to implement a modified Winston-Lutz test for IGRT accuracy required for optimal treatment delivery. This system is also used for TG-142 quality assurance compliance for isocenter and laser alignment tests.
End-to-end positioning accuracy from planning CT to clinical IGRT treatment setup is achieved with 0.1 mm precision and is accomplished in minutes. The test is performed by aligning the QUASAR™ IsoCenter Cube phantom using the cross-hairs etched on each side with your IGRT system, then capturing radiation field exposures with film or EPID. The test results are used to determine end-to-end isocenter accuracy for frameless SRS, SBRT, and IGRT treatments.
Features
Workflow Efficiency
Quick and simple Winston-Lutz testing for IGRT accuracy. Integrate the QUASAR™ IsoCentre Cube into your daily QA routine for simple verification of your laser, isocentre and couch adjustment alignment.
Simple Design
The QUASAR™ IsoCenter Cube Phantom features a seamless 5 cm³ acrylic unibody enclosure containing one 6 mm diameter Tungsten Sphere located at the central axis of the main cross-hairs and used for measuring isocenter accuracy.
Image Match Verification
Displayed on three faces of the QUASAR™ IsoCenter Cube Phantom are 3 mm offset laser alignment marks, which are used to easily verify the couch position adjustment accuracy after image matching.
Isocentre Verification
User confidence is increased with the verification of your isocentre alignment accuracy. Report detailed statistics including left/right, superior/inferior and anterior/posterior errors at configurable gantry, collimator and table angles.