noise:<45dB
flow rate: 0~5.0L/min
pressure error: ± 2cmH2O
pressure adjustment stride: 1cmH2O
Data statistics: statistics of flow and effusion
Built-in power: standby time is greater than 8 hours
pressure adjustment range:-1cmH2O ~-100cmH2O
weight: 1.5kg
Length × Width × Height: 225mm × 90mm × 215mm
safety protection: leakage, blockage, full bottle, dumping, low battery alarm
01
automatic super negative pressure release function
If the intrathoracic pressure is -15 cmH2O or higher, and the difference between the set pressure and the intrathoracic pressure is -5 cmH2O and lasts for 10 seconds or more, the high negative pressure will be automatically released.
02
Multi-cavity pathway
is used to measure intrathoracic pressure and fluid tank pressure. When exudate accumulates in the drainage tube and a pressure difference occurs between the drainage tube and the measuring tube, the system will perform pressure compensation to correct the pressure difference.
Anti-overflow protection
There is an anti-overflow protection alarm device in this machine, which will automatically stop working and alarm when the liquid level reaches the warning line.
prevent the liquid from flowing back to the main machine and cause damage to the machine to avoid secondary pollution.
Air leakage alarm
When a liquid leak occurs, the device will automatically stop running and alarm
Other security protection
dumping, blockage, full bottle, low battery alarm, etc.