It is mainly used for in vitro quantitative detection of serum amyloid A (SAA) content in human serum and plasma.
Intended Use
Serum amyloid A is an acute-phase protein whose concentrations in the blood can increase sharply within hours during inflammatory, infectious, and non-infectious diseases. Can be increased to 1000 times the original concentration. Serum amyloid A is associated with high-density lipoprotein (HDL), which regulates HDL metabolism during inflammation.
As with C-reactive protein (CRP) (similar), the detection of serum amyloid A is useful in diagnosing inflammation, assessing its activity, monitoring its activity, and treating it.
Advantage
No cold chain required.
Single packaging.
Compact and convenient.
Principle
Serum Amyloid A Test Kit (Chemiluminescence Immunoassay) adopts the chemiluminescence competition method.
Main Components
25 Lyophilization sphere reagents
Sample Requirements
1. The serum is recommended to be collected by ordinary serum collection tubes or tubes with procoagulant. If it cannot be used immediately, it is recommended to be stored at 2-8℃ and the test should be completed within 24 hours. If the test cannot be performed within 24 hours, the sample needs to be frozen at -20℃ for 1 month.
2. Samples containing sediment or flocculent material should be centrifuged (4000rpm, 10min) and then tested.
3. Samples must be returned to room temperature (15-30℃) before testing. Frozen samples should be completely thawed, rewarmed and mixed well before use. Do not repeatedly freeze and thaw.
4. Heat inactivated samples, hemolysis and other abnormal samples should be discarded.