It is mainly used to quantify in vitro the content of Beta-Human Chorionic Gonadotropin(β-hCG) in human serum and plasma.
Applicable Devices
Beta-Human Chorionic Gonadotropin(β-hCG) Test Kit (Chemiluminescence Immunoassay) used for Poclight C5000 chemiluminescence immunoassay analyzer
Sample Requirements
1. The serum is recommended to be collected by ordinary serum collection tubes or tubes with procoagulant. If it cannot be used immediately, it is recommended to be stored at 2-8℃ and the test should be completed within 24 hours. If the test cannot be performed within 24 hours, the sample needs to be frozen at -20℃ for 1 month.
2. Samples containing sediment or flocculent material should be centrifuged (4000rpm, 10min) and then tested.
3. Samples must be returned to room temperature (15-30℃) before testing. Frozen samples should be completely thawed, rewarmed and mixed well before use. Do not repeatedly freeze and thaw.
4. Heat inactivated samples, hemolysis and other abnormal samples should be discarded.