Dental implant stability meter (tapping-motion) for measuring stability of an implant without disconnecting the healing abutment.
Features
1)EASY & FAST
It is safe by tapping directly into the Healing Abutment,Measurement is faster (within 3 seconds) and easier compared to the RFA Type.
2)EFFICIENT
It can be measured for natural teeth or implant prosthesis by tapping directly on the measuring body.As a result, it is more utilized than the RFA type not only in determining the loading time but also in early diagnosis of inflammation around the implant.
3)SAFE
Compared to existing hitting products, the number of hits has been drastically reduced (16 times → 6 times), and if the fluctuation level is less than 59, it is safe by automatically stopping after hitting twice.
1)Measurement method
Turn on AnyCheck by pressing the switch button until it turns on.
Slightly contact the tip of the tapping rod to the upper part of the healing abutment. Be cautious not to push the healing abutment.
Maintain the contact angle between 0 to 30 degrees.
Gently press the START button while holding the device stable
Check the measured value displayed on the LCD screen.
Range of the measured value is between 30-85. The smaller the measured value, the weaker the degree of osseointegration.
2)Check the measured value
Measured values 1 to 99 are displayed. (1 to 59 red, 60 to 64 orange, 65 to 99 green)
The lower the measurement value, the greater the implant fluctuation
Measured values (IST scale) differ depending on the height of the healing abutment.