This kit is a surgical tool to remove failed implants while minimizing bone damage.
Features
Minimize bone loss while removing implant
No need to use a trephine drill
Use Fixture remover screw, Fixture remover and Torque wrench
Prevents loss of cortical bone thread line
FR Kit only rotates failed implant to remove
Fixture with identical diameter as removed one can be implanted with remaining bone thread line
Applicable to any type of implants with a screw hole
FR Kit is equipped with fixture remover screw of various sizes.
Available from M1.4 to M2.5 of screw hole size
Components
1.Fixture Remover Screw
2.FR Screw Hex Driver
3.Fixture Remover
4.Torque Control Device
5.Fixture Remover
Procedures
Failed Implant
Completely remove the prosthesis of the implant that needs to be removed.
Fixture Remover Screw
Select an appropriate size of Fixture Remover Screw.
Connect the Fixture Remover Screw to the FR Screw Hex Driver and tighten it clockwise.
Fixture Remover
Select an appropriate size of Fixture Remover.
Connect the selected Fixture Remover and tighten it counterclockwise until it is completely tightened to the Fixture Remover Screw that is fixed in the implant.
Using a Torque Wrench, further apply torque to the Fixture Remover(counter clockwise) until the implant is finally unscrewed from the site.
Clear