This kit is a surgical tool to remove fractured screws.
Features
Remove a fractured screw without damaging the fixture
Perfect Guide and Conical, Internal, External Guide act as the guide when using the reverse drill.
Make a precision hole to connect the screw remover without damaging internal screw thread of implant.
Capable to remove fractured screws from different implant system.
SR Kit is equipped with reverse drill and guide of various sizes.
Available from M1.8 to M2.5 of screw hole size
Components
1.Claw
2.Reverse Drill
3.Screw Remover
4.Shank Driver
5-1.Perfect Guide & PG Hand Driver
5-2.Perfect Guide & PG Hand Driver
6-1.Conical, Internal, External Guide
6-2.Conical, Internal, External Guide
6-3.Conical, Internal, External Guide
7.SR Driver Holder
Procedures
Failed Implant
Completely remove the prosthesis of the implant that needs to be removed.
Perfect Guide
Select an appropriate size of Perfect Guide.
By using the PG Hand Driver, tighten the Perfect Guide To the fixture in a clockwise direction until it touches the screw inside the fixture.
Reverse Drill
Select an appropriate size of Reverse Drill and connect it to the contra angle of the surgical engine.
Insert the Reverse Drill into the Perfect Guide attached on the fixture.
By drilling counter clockwise at a speed of approximately 2,000rpm, drill 1~2mm into the surface of the fractured screw.
Check the marking line(in 1mm intervals) of the drill body while drilling with adequate irrigation.