3-in-1 pre-wired electrode array
Pre-measured sensor position
Radiolucent material nearly invisible on x-rays
One touch application ideal for use in delivery room
Larger surface area for enhanced adherence
Engineered for back application, allowing chest to remain exposed
Multi-colored wires for easier identification
Not made with natural rubber latex or plasticizer DEHP
Individually packaged
Heart rate is a major indicator for decision-making during delivery room assessment of a newborn. The NeoHeart ECG Pad is the ideal answer to the latest NRP recommendations. It’s designed for short-term use during labor and delivery. The pre-measured electrode placement allows for one touch application, rather than placing three individual leads. Unlike our NeoLead electrodes, NeoHeart does not have hydrocolloid so there is no warming time. And there’s no need to wipe the baby clean first. NeoHeart has a larger surface area with a stronger adhesive for better adherence over vernix.
Application
Step 1 Wipe and remove any excess moisture from where NeoHeart will be applied.
Step 2 Peel and discard liner. Apply immediately after liner is removed
Step 3 Place NeoHeart on patient’s back and connect to ECG monitor cable. The pad should be oriented so the point of the heart is pointing toward lower extremities.
Step 4 Confirm quality of ECG tracing.
Note: if QRS is stable but appears small or dampened, consider changing to a different lead on the monitor or increasing the size of the waveform. If small or dampened waveform persists, application to the patient’s chest may result in a stronger waveform.