Features & Settings
CONTROLLED MODE
Safe treatment programming for patient’s use at home.
The physician sets:
Number of stimulations (1—99)
Stimulation intensity and mode (anodal, cathodal, sham)
The minimum time interval between two consecutive stimulations (0—168 hours)
HDCstim delivers the tDCS treatment only following the programmed settings
FREE MODE
Healthcare professionals and researchers can use HDCstim in a free mode according to their needs.
Waveform: one channel, two channels, sham
Sham time: 0—100%
Stimulation amplitude: 200—2000 µA
Ramp duration: 5—30 seconds
Intensity: up to 2 mA per each channel
Duration: 1—40 minutes
FEATURES
Portability: Small size, 12cm x 8cm x 3cm
Two-channel stimulation: Anodal only
Maximum output: 30 Volt - 2 mA (4 mA with HDCadpt accessory)
Treatment monitoring: Impedance check and stimulation failures
Double blind: Sham mode not distinguishable by user
Impedance check resolution: 1kΩ
Output current stability monitoring: Real time - 1 second
Internal power supply: 2 x 1.5 Volt type AA
Memory: Maximum 99 stimulation reports
Medical device classification: CE Class IIa, CMDCAS Class IIa, GMP class III
Home use: Compliant to IEC 60601-1-11:2011
Intellectual property: EP2328655 US9,026,216