◆1-8 Channel Waveform Display
◆7 Leads ECG Waveform Display; ST Segment Analysis
◆Defibrillation and ESU Protection
◆Venous Punch
Product Specification
♦ECG
ECG cable: 5 Leads ECG Cable
Lead selection:I,ïï, ni aVR, aVL, aVF/ V
Gain (mm mV): 1/4,1/2,1, 2,4, Manual Automatic
Sweep speed (mm sec): 6.25,12.5, 25, 50
Heart rate range: 30~300bpm
Heart rate accuracv:=lbpm or ±\ %
ST segment analysis (Diagnostic mode):
a. ST range: -2.0mV—i-2.0mV
b. ST accuracy:±0.02mV
Monitoring mode: Diagnostic 0.05 ~ 100Hz
Monitoring 0.5 ~ 40 Hz
Operating 1.0 ~ 25 Hz
♦Non-Invashe Blood Pressure
Measurement type: adult, pédiatrie, neonatal
Measurement range:
Adult: Systolic 40 ~ 270mmHg
Diastolic 20 - 200mmHg
Mean 30 - 220mmHg
Pédiatrie: Systolic 40 - 180mmHg
Diastolic 15 - 160mmHg
Mean 20 - 170mmHg
Neonatal: - Systolic 30 ~ HOmmHg
Diastolic 15 - lOOmmHg
Mean 20 ~ 11 SmmHg
Accuracv: Meets ANSIAAMI SplO standards
Over pressure protection
♦Temperature
2 Channels
Measurement range: 25 ~ 45 °C
Accuracv :±0.2 °C
♦Options
EtC02
Measurement range: (MSOmmHg
Accuracv: ±2 MmHg (0~40)mmHg: ±5%(41~70)mmHg
±8% (71~100)mmHg:=10%(101~150)mmHg
Response time: <3 seconds