This section provides information on the operating environment, physical controls, and dimensions of the DH-2000. It also provides pinout information for the connector.
Operating Environment
The following table provides information on optimizing the operating environment of your DH2000.
Moisture 5-95% without condensation at 40°C
Ventilation Unit should be situated so that its location or position does not interfere with proper ventilation.
Temperature 5°C – 35°C. Unit should be situated away from any device that emits excessive heat.
Object and Liquid Entry Unit should be positioned so that objects do not fall on top of the unit. Additionally, ensure that no liquids are spilled into the enclosure through openings.
Description
Deuterium On/Off
Press the blue DEUTERIUM button to turn the Deuterium lamp on or off. The bulb requires a warm-up time of 20 seconds before the Deuterium lamp is illuminated. You must allow the bulb to warm up to receive accurate data from the lamp. LED located under button lights green upon successful illumination or red to indicate lamp malfunction.
Mechanical Protection – SMA Connector
Covered to protect users from unintentionally looking directly at the beam of light. Connect the fiber cable to the DH-2000 BEFORE turning the lamp on to avoid unnecessary exposure to UV radiation. Always wear proper eye protection when using the DH-2000 lamp.
Halogen On/Off
Press the red HALOGEN button to turn the Halogen lamp on or off. LED located under button lights green upon successful illumination or red to indicate lamp malfunction
Shutter Mode Switch
Sets the operational mode of the shutter.