Work smarter with the Model 5390 Tandem PID/XSD, which combines the power of the 4430 Photoionization Detector (PID) and the 5360 Halogen Specific Detector (XSD) to simultaneously detect halogen and aromatic compounds.
Overview
This patented tandem detector offers the production of two gas chromatograph (GC) detectors, the 4430 PID and 5360 XSD, in one efficient package. Its unique design eliminates transfer lines and uses only one detector port. The model 5390’s simplified design improves reliability and performance, and reduces costs. Detectors can be used individually if desired.
5390 Tandem PID/XSD Features:
Unique design eliminates need for transfer lines and minimizes dead volume
Patented tandem detector design uses only one GC detector port
Exclusive Lampsaver™ circuit can be programmed to turn off PID lamp after a specified period to extend lamp life
High sensitivity with selective detection of halogenated compounds
Low maintenance and increased stability and reliability
Unique jet design minimizes peak tailing due to unswept dead volumes or transfer lines
Specifications
Dynamic Range>105; Linear Range > 104
Detectivity 104
Reactor Operating Temperature1,000 – 1,100 ˚C
Gas Requirements/ Flow RateUltra high purity air (or oxygen) 10-30 mL/min
Power Requirements90–260 VAC (±10% VAC)
47–63 Hz, 200W
Detector Weigh 0.8 lb (0.36 kg)
Signal Output 0–1V or 0–10 V
Dynamic Range - > 104
Sensitivity - <40 pg benzene
Lamp Current - 0 – 1.35 mA in 9.15 mA steps
Lampsaver Time - 0.5 – 2 hr, reset by external contact
Gas Requirements - He 99.999%
Power Requirements - 105 – 125 VAC, , 25 VA
210–240 VAC, , 25 VA
Maximum PID Operating Temperature - 250 ˚C