The Model 5360 XSD™ offers superior selectivity for detection of halogen-containing compounds.
Overview
Installed in the standard port of a gas chromatograph (GC), the Model 5360 XSD™ detects halogen-containing compounds eluting from the GC’s capillary column. This instrument is unique in the fact that it doesn’t contain radioactive sources or use organic solvents like other halogen selective detectors. This eliminates the need for catalyst tubes, solvents, resin cartridges, pumps, and transfer lines.
The 5360A XSD™ option offers a heated base assembly which provides improved baseline stability and easy column installation. Both the 5360 and 5360A models can be installed as stand-alone detectors or used in tandem with the Model 4430 Photoionization Detector (PID).
XSD™ Features:
High halogen selectivity vs. hydrocarbon simplifies analyses and minimizes (or eliminates) the need for sample preparation
High detector sensitivity permits very low-level selective analysis of halogen-containing (Cl, Br, F) compounds
Unique jet design minimizes peak tailing due to unswept dead volumes
Requires only air to operate
Designed to operate on most GC makes and models
No routine maintenance required
Enhanced venting option diverts solvent before entry to the reactor
Specifications
Dynamic Range - >105
Linear Range - >104
Detectivity - 104
Reactor Operating Temperature - 900-1100 ˚C in 100 ˚C increments
Flow Rate - 20-30 mL/min air
Communications Signal Output - 0-1 V or 1-10V
Gas Requirements - Air 20-30 mL/min (ultrahigh purity)
Power Requirements - 90-260 (±10%) VAC, 47-63 Hz, 200W
Controller Dimensions - 21.2 cm H x 12.8 cm W x 30.8 cm D