Pulsar is a compact platform for anterior segment surgery that implements the patented Minimal Stress technology: the system capable of controlling the power delivered to the handpiece in order to maintain the exact tip stroke set.
Minimal Stress technology guarantees the safest cataract extraction, optimizing energy at the lowest efficient level.
Its Fluidics is based on an high performance peristaltic pump that ensures maximum chamber stability for an optimal patient safety.
Pulsar is equipped with a Dual Linear Pedal, designed for complete control, flexibility and comfort, offering a wide range of customization.
Pulsar offers a wide range of reusable and disposable accessories such as reusable and disposable I/A tubing set, reusable and disposable phaco tips for incision from 3.2 to 1.8 mm and four crystal titanium phaco handpiece.
EQUIPMENT CONTROL
Footpedal
single linear, single linear left, double linear, double liniear left, MMP, single IVP
IRRIGATION
Irrigating pressure
regulated by the height of the electric I.V. pole
Irrigation control
footpedal and arrow keys on frontal panel
ASPIRATION
Aspiration pump type
Peristaltic with single-use or reusable tubing
Vacuum level range programmable from 5 to 500mmHg (step 5mmHg)
Flow rate range
programmable from 2 to 50cc/min (step 2cc/min)
Rise time (in us mode)
25 levels, programmable from 2 to 50cc/min (step 2cc/min)
Control
panel or linear vacuum and flow control by the footpedal
Safety system
ACS3 (Anterior Chamber Surge Suppression System)
VITRECTOMY
Cutter type
pneumatic guillotine cutter