laser light stimulates the body’s own healing response to lower eye pressure
effective as primary, adjunctive or repeat therapy
safe, office-based procedure
reimbursed by most insurance policies
Key features
Optotek has spent extra time and effort to ensure highest reliability and safety of the system. The OptoSLT M incorporates a PPS™ technology for delivering stable output energy far exceeding current industry standards. Enhanced accuracy and stability of energy provide the highest degree of safety and assure effective and efficient eye treatments.
e-SlitLight™
HIGH POWER LED ILLUMINATION SYSTEM
FES™ - Fine Energy Setting
FOR UNMATCHED PRECISION
PPS™ - Pulse to Pulse Stability
FOR ADVANCED SAFETY
Treatment at 3 Hz
FOR HIGHER PERFORMANCE
Easy-to-move System
MECHANICAL PRECISION AND OPTICAL QUALITY
Technical specifications:
Laser
Energy 0,2 to 2,6 mJ per pulse in 33 steps
Spot Size 400 μm
Cone Angle < 3°
Pulse Rate - Up t 3 Hz
Burst Mode Single pulse
Pulse Duration 4 ns (typical)
Aiming Beam Red diode; 650 nm; continuously adjustable intensity(635, 670 nm optionally)
Laser Trigger Joystick (footswitch optionally)
Cooling System - Air convection
Slit Lamp
Illumination Source White LED
Magnification 10x, 16x, 25x (6x, 10x, 16x, 25x, 40x optionally)
Table
Type Central mounted
Dimensions 600 x 480 mm
System dimensions
SLT Laser Head Dimensions 240 x 150 x 60 mm
Total System Height 695 – 730 mm
Electrical Requirements
Power Supply 100 – 240 V, 50 – 60 Hz, 1,1 A MAX