The BiPAP AVAPS device is intended to provide non-invasive ventilatory support to treat adult and pediatric (> 7 years of age and > 40 Ibs) patients with obstructive Sleep Apnea (OSA) and Respiratory Impairment. The device may be used in the hospital or home.
Specifications
Humidification Heated humidification: fixed, adaptive
Device controls LCD, control dial/push button
Filters Reusable pollen; disposable ultra-fine
Warranty 2 years
Breath per minute 1 to 30 (increment is 1 breath)
AVAPS
Target volume: 200 to 1500 ml per breath (increment is 10 ml)
IPAP Max : up to 30 cm H₂O
Min. IPAP: EPAP plus 2 cm H₂O to 30 cm H₂O (Minimum Pressure support is 2 cm H₂O)
Inspiration time 0.5 to 3.00 seconds (increment is 0.1 seconds)
Ramp time 0 (off) to 45 minutes (increment is 5 minutes)
Humidification Heated humidification, fixed, adaptive (standard 15mm or 22mm or 15mm heated tube)
Rise time 1 to 6 (100 to 600 ms)
Automated Airway Management (AAM) (selected countries) EPAP Max : 4 to 25 cm H₂O EPAP Min : 4 to 25 cm H₂O Pressure Support: 0 to 26 cm H₂O
Dimensions and weight
Dimensions
15.7 x 19.3 x 8.4 cm (w/o humidifier) ; 29.7 x 19.3 x 8.4 cm (w humidifier)
Weight
1.33 kg/2.94 lbs (w/o humid); 1.98 kg/4.37 lbs (w humid); Incl power supply
Electrical
Voltage
100 – 240 VAC
Amperage
2.0-1.0 A
Frequency
50/60 Hz