Pigeon Medical air-oxygen blender is a device used by medical personnel to control the concentration and flow of oxygen for the rescue of patients with hypoxic problems while avoiding the side effects of pure oxygen inhalation. It provides the ability to mixed air and oxygen in a safe, easy, and controlled manner, supporting the use of a ventilator, HFNC machine, infant resuscitator, infant radiant warmer, CPAP, especially suitable for those who have high requirement for oxygen concentration.
Working Principle of AD3000-SPDF Air Oxygen Blender
It is also equipped with a previse balance device to output stable oxygen concentration from 21% to 100% optionally. The FiO2 can be controlled by adjusting the proportion of air and oxygen flow.
Features of AD3000-SPDF Air Oxygen Blender
Selection of flow meter from 0.1-6LPM/ 1-15LPM/ 6 to 60LPM each side
Can match with humification bottle
Adjust oxygen concentration level from 21% to 100%
Specification of AD3000-SPDF Air Oxygen Blender
Internal Filter (Air and Oxygen Inlet) - 80~100μm
Oxygen Concentration Adjustment Range - 21% - 100%
Accuracy of Oxygen Concentration - ±5% FS
Gas Supply Pressure - 0.35±0.02 MPa(50±3 PSIG)
Outlet Flow Range - 6 LPM (AT 21% Oxygen and 100% Oxygen) to 120 LPM (AT 60% Oxygen)
Optimal Working Pressure - 0.35 MPa(50 PSIG)
Maximum Allowable Pressure - 0.7 Mpa(100 PSIG)
Temperature Range - 10~50℃
Ambient Storage Conditions Temperature Range - -20~60℃
Relative Humidity Range - 0 - 95% Non-Condensing
Practical Application of AD3000-SPDF Air Oxygen Blender
Bubble CPAP
HFNC
Ventilator gas supply
Routine adult oxygen therapy
Anesthesia Machine
Incubator