The Powerex FirstCall™ medical gas alarm systems monitor the status of medical gases in Category 1 healthcare facilities. The alarm systems are ETL listed to UL 1069 and comply with the latest edition of NFPA 99. All alarm systems provide audible and visible indications of NFPA 99 medical gas alarm conditions.
10” high-resolution touchscreen HMI with 7” VE option available
Ethernet connectivity using BACnet over IP protocol to building management system – standard
Customizable emergency instructions in the event of an alarm
Area alarm up to 8 gases
Master alarm 23, 39, and 55 input signals
Combination alarm up to 6 gases, 23, 39, or 55 input signals
Designed and manufactured in the USA
5 year warranty on parts, 2 year warranty on labor
Powerex UL ETL Listing (UL 1069)
Powerex CSA ETL Listing (CAN/CSA C22.2 No. 205)
Features
Built by Powerex in Harrison, OH
5 year warranty on parts, 2 year warranty on labor
10” high-resolution touchscreen HMI with 7” VE option available
Area alarm up to 8 gases
4-20mA signal transducers, DISS demand-check fittings, and gas risers included with Area Alarm
Transducers are able to be remotely installed up to 5,000 ft away
Master alarm 23, 39, and 55 input signals
Combination alarm up to 6 gases, 23, 39, or 55 input signals
Ethernet connectivity using BACnet over IP protocol to building management system – standard
Alarm signal status can also be viewed on a standard computer connected to the facility’s local network via Ethernet
Customizable emergency instructions in the event of an alarm