Three-step defibrillation process
Two-button operation
Extensive voice and visual prompts for the operator
Biphasic energy output
Lock-out protection to prevent inadvertent defibrillation
Specification
Category - Specification
Output - Biphasic Truncated exponential
Analysis Time - 9 sec.
The maximum time from the initiation ofrhythm analysis to readiness for discharge with a new battery
- Less than 30 seconds
The maximum time from the initiation ofrhythm analysis to readiness for discharge after 6 shocks - Less than 35 seconds
The maximum time from the initially switching power to readiness for discharge - Less than 25 seconds
Audible Prompts - 20 audible prompts
Visual Prompts - LED prompts
Controls - Two buttons- On/Off. Shock
Output Energy Accuracy - ±15% into any impedance from25 to 175 Ω
The Maximum Voltage - 120±50V
Output disabled when the patient impedance is outside limits - 20Ω to 200Ω
Battery
Part No - CR123A-4*2
Non-rechargeable - 12V d c. 3 .0Ah
Capacity - 100 discharges at 200 Joules or 120 discharges
150Joues
Shelf Life (25℃±15℃) - 10 years(5 years storage +5 years standby)
5 years standby(after installation)
Physical
Operating Temperature - O℃ to 40℃
Operating Humidity - Relative humidity between 30% and 95% (non-
condensing)
Storage Temperature(without battery) - 20℃ to 55℃
Storage Humidity (without battery) - Up to 93% (non-condensing)