Power requirement : AC 220 V, 50 Hz,
Heater output : 650 Watt
Halogen Lamp : 220 Volt
Temperature Control Range : 26,0 oC – 37,0 oC ( air and skin temperature)
Temperature Accuracy : 0,1oC – 0,3oC
Room Humidity : 25 % – 75 %
APGAR Timer : adjustable reminder tone Up to 24 hours set time
Room temperature : 18,0 oC – 30,0 oC
Room air flow : max 0,3 m/s
Skin censor accuracy : max 0,3oC
Skin and controller temperature deviation : max 0,5 oC
Temperature control mode : skin or air
Heater indicator display : on/off
Heater warm up time : max 30 min
Setting temperature fluctuation : max 0,3 oC
Heater box rotation angle : 0 – 135 o (two ways with lock buton)