AC Power supply: 100-240 V~
Frequency: 47 - 63 Hz
AC Input power: 1.0A(100V~) / 0.6A(240V~)
Battery: Up to 4 hours autonomy.
Built-in charger.
Environmental conditions:
Storage temperature: -20ºC to 40ºC
Operating temperature: 0ºC to 40ºC
Altitude: -300 m to 2.000 m
Alarms: Optical and acoustic for all parameters.
Three priority levels; clinical, technical and procedural.
Trends: 72 hours for all parameters, graphical and tabular display.
Communication module for data output.
Connectors: serial DB9 (RS232), RJ45, USB
ECG
Leads: I, II, III, aVL, aVF, aVR, Vx
Waveforms on screen: 1, 2 or 7
Resolution: 1 bpm
Accuracy: ± 2% or ± 2bpm
Freeze and memory: 12 seconds of ECG strip for on screen measurement
CMRR: > 115 dB in monitor mode > 90 dB in diagnostic mode
Signal features:
Diagnostic mode: 0,05 to 100 Hz
Monitor mode: 0,5 to 40 Hz
O.R. mode: 0,5 to 20 Hz
Programmable gain: 1 to 8 mV
Pacemaker detection and rejection: Pulse duration: 0,1 to 2 ms
Pulse amplitude: ± 2 to ±700 mV
Patient safety: CF type
NIBP
Time: typical 25 to 45 seconds maximum 120s (adult/child) 90s (neonate)
Modes: Manual Automatic: 1 to 120 min. (1 min. interval)
Patient safety: BF type