AC Power supply: 100-240 V~
Frequency: 47 - 63 Hz
AC Input power: 1.0A(100V~) / 0.6A(240V~)
Built-in charger.
Environmental conditions:
Storage temperature: 20QC to 40QC
Operating temperature: 0QC to 40QC
Altitude: -300 - m to 2.000 m
Alarms:
Optical and acoustic for ail parameters.
Three priority levels; clinical, technical and procedural.
Trends:
72 hours for ail parameters,
graphical and tabular display.
Communication module for data output.
ECG
Leads: I, - II, - III
Resolution: 1 bpm
Accuracy: ±2% - or±2bpm
Freeze and memory:
12 seconds of ECG strip for on screen measurement
CMRR:>115dB in monitor mode
> 90 dB in diagnostic mode
Connectors: serial DB9 (RS232), RJ45, USB
Signal features:
Diagnostic mode: 0,05 to 100 Hz
Monitor mode: 0,5 to 40 Hz
O.R. mode: 0,5 to 20 Hz
NIBP
Time: typical 25 - to 45 - seconds
maximum 120s
Automatic: 1 to 120 min. (1 min. interval)
Patient safety: BF type
Programmable gain: 1 to 8 mV