4-channel filter photometer
Three channels for UV and one for thevisible range
Light Source - LED with long lifetime
Wavelength - 340 nm and 380 nm – 730 nm
Wavelength Selection
• Positions #1-3: fixed at 340 nm
• Position automatically via
9-position filter wheel: 5 standard interference filters:
405 nm, 492 nm, 546 nm, 578 nm, and 623 nm;
4 positions for optional filters of choice
Photometric Range - 0 – 2.5 A
Cuvette System
Normal standard cuvettes (macro or semi-micro, disposable or special optical glass)
Temperature Control
Measuring cells are temperature
controlled by electrical thermostat at 37 °C
Operator Interface
Touchscreen for direct functions and alphanumerical inputs
Data Presentation
Graphic display: white characters or symbols, blue backlight, dimension 240 * 128 dots
Integrated Printer
Thermal printer, 24 characters per line
Memory -
General operating software can be updated by PC
Reagent open system with capacity for up to 231 programmable methods
Import of data by touchscreen or PC
Up to 50 non-linear calibration curves with max 20 sets of points can be stored
Data Logging
Up to 1000 results can be saved in the memory
Signal Port
Serial port for connection to an external printer and/or PC
RS232 connector for barcode reader for an easy management of sample identities
Quality control -
Up to 50 methods can be controlled with two control sera
Levey Jennings plot
Measuring Time -
Kinetic: variable from 3 – 19 deltas, time per delta 3 – 255 sec
Fixed time: variable from 0 – 1800 sec
4-Channel Mode:
Three positions at 340 nm for Kinetics with factor
One position for end point with factor in the visible range