The original IITC Hot/Cold Plate Analgesia Meter available with ramping temperature from -10 deg. C to 70 deg. C. IITC Life Science developed the first commercially available test apparatus in the world which is capable of truly assessing pharmacological modulation of thermonociception.
Temperature Range: -10 C to 70 C
Ramping temperature for threshold & latency results
Rapid increase or decrease in temperature
Supplied with a clear animal enclosure
Plate size 4” x 8”
Precise Programmable digital control
Print out of data
Temperature stability is .1 deg. C.
Footswitch – RS232
Software – USB – (optional)
IITC’s Hot/Cold Plate Analgesia Meter widens the range of analgesia research to new levels. The unit is microprocessor controlled and can be used for both incremental heating and cooling from any starting temperature from -10 C to 70 C in 0.1 C increments. The microprocessor monitors and controls the rate of heating/cooling which can be set from 1 to 10 C per minute. The front panel display shows all test results including length of test in .01 sec. increments, starting temperature and stopping temperature. The plate is uniformly heated/cooled at an adjustable rate and has automatic upper/lower cut-off limits. All parameters are entered via front panel keypad. i.e. rate of temperature increase/decrease, maximum/minimum temperature, date/time and animal number. The IITC system can also double as a constant temperature hot plate analgesia meter.
Upon reaction of either the hind paw or fore paw (whichever is chosen) a foot switch is depressed and the heating/cooling of the unit is reversed to the standby temperature within minutes.