Basic humidification system for Ventilator. This device controls the temperature of the electrical heating plate to raise the temperature of the distilled water in the chamber and adjust the respiratory gas supplied from the mechanical ventilator to the patient at appropriate temperature and humidity level.
Features
Safety feature
An anodized finish on heater plate is non-toxic. The heater plate will function with the insertion of the humidifier chamber. In addition, if there is no chamber, it automatically power off. There is a built-in thermal protection against over heated
Light weight and simple
The total of the device is only 0.82kg. User interface is simple and easy to handle for everyone.
With the Quick Release, it is easy to connect/disconnect the chamber to the device.
Easy drain design and durability
The heater plate is anodized Aluminum.
Exterior anodic coatings provide good stability to ultraviolet rays, and do not chip or peel easily.
Specification
Temperature setting 34 ~ 70°C (Heating plate)
Safety feature Operation of heating plate: SH-2 doesn’t work unless humidifying chamber is installed to the device.
Humidification output >33mg/L at 46°C Target (Level 4),
>12mg/L at 42°C Target (Level 3)
Rated voltage 220V~
Rated frequency 50/60Hz
Rated power 160VA
Electrical protection BF, Class I
Dimensions 138x170x87 mm (LxWxH)
Weight 0.82kg (without accessories)
Electrical Protection BF, Class I
Degree of Protection against Ingress of water IP22