This high flow pulsating pump is used to generate periodic pulsating fluids, simulate human heart beating, and dynamically simulate arterial and venous blood vessels, including periodic pulsating circulation, liquid temperature control, and pressure monitoring. It can be used for relevant experimental research by medical researchers. The liquid circulation module can achieve liquid heating, temperature control, and circulation. The control module can control the amplitude, frequency, flow rate, pressure, and temperature of the driving motor.
Equipment parameters:
1. Instrument model: SHS-1800
2. Voltage/Frequency: 90-240VAC/50-60Hz
3. Pulse frequency: 20-180 Beat/min
4. Cardiac output: 7 liters/ min@Max (72 Beat/min)
5. Pressure range: 0-250mmHg
6. Pressure accuracy: ± 2mmHg
7. Temperature control accuracy: ± 0.5 ℃
8. Maximum power: 800W
9. Instrument appearance dimensions: 290mm × 165mm × 290mm
10. Gross weight of the instrument: 10kg
Software functions:
1. The various functional areas of the operation interface are clear and can directly reflect the real-time status of the pump:
2. Easy to set up and operate, frequency and traffic can be operated simply by clicking on it;
3. The flow rate can adjust the output and input size of the pump separately;
4. Two sensors collect data and can observe data from two different locations:
5. The pressure difference function can analyze the pressure difference curve at different positions:
6. Data collection function: It can collect raw data on the pressure, pressure difference, and equipment flow of two channels separately, and generate txt text format for data traceability and analysis;