ENVITEC OOM202 Oxygen Sensor O2 Cell
Part Number: OOM202
Specifications:
Measuring range: - 0-100%
Output signal in the air: - 13-16mV
Circuit interface: - 3Pin(Molex)
Precision and repeatability: - <1% oxygen by volume, after calibration in 100% oxygen
Linearity error: - <3% relative error
Response time: - <12 seconds, relative to 90% relative value
Zero drift: - <200 u V placed in 100% ammonia for 5 minutes
Cross interference: - <0.5% volume oxygen for:
- 10% co2, in balanced ammonia
- 80% N20, in balanced ammonia
- 7.5% halothane, in balanced ammonia
- 7.5% isoflurane, in balanced ammonia
- 7.5% enflurane, in balanced ammonia
- 9% enflurane, in equilibrium nitrogen
- 20% desflurane, in balanced nitrogen
Influence of temperature: - -0.03% relative temperature, 25℃
The effects of stress: - Proportional to oxygen partial pressure
Mechanical shock effects: - <1% relative error, after falling from a height of 1 meter
Operating temperature: - 0℃-50℃
Temperature compensation: - Built-in thermistor compensation
Temperature effect - 25℃-40℃, 3% relative error
Compensation (steady state): - Between 0℃-50℃, relative error of 8%
Working humidity: - 0-99% RH (several hours in condensable air)
Storage temperature: - -20℃- 50℃
Recommended storage temperature: - 5℃ ~ 15℃
Recommended load: - ≥ 10Ω
Preheat time: - <30 minutes, after replacing the sensor
Rated sensor life: - ≥ 1000 000% volume oxygen hour
Weight: - About 28G