Nominal sensor lifetime ≥ 500 000 % volume oxygen hours
Output in ambient air 14 mV ... 20.7 mV (Dual Cathode), load 600 Ohms
Electrical interface Gold plated slip rings
Accuracy meets ISO 80601-2-55 requirements
Repeatability < 1 % volume O2 at constant temperature and pressure
Linearity error 3 % relative
Response time < 12 s to 90 % of final value
Zero offset voltage < 200 µV in 100 % nitrogen, applied for 5 min
Cross interference meets ISO 80601-2-55 requirements
Influence of humidity -0.03 % rel. per % RH at 25 °C
Pressure range 0.6 bar ... 2 bar (ppO2 0 ... 1250 mbar O2 )
Influence of pressure proportional to change in oxygen partial pressure
Influence of mechanical shock < 1 % relative after a fall from 1 m
Operating temperature 0 °C ... +50 °C
Temperature compensation - no temperature compensation
Operating humidity 0 % ... 99 % RH non-condensing
Long term output drift < 1 % volume oxygen per month typically < -15 % relative over lifetime
Storage temperature -20 °C ... +50 °C
Recommended storage +5 °C … +15 °C
Recommended load ≥ 10 kOhms
Warm-up time < 30 minutes, after replacement of sensor
Weight approximately 28 grams
Use the advantages:
• Compliant with European MDD (CE certification)
• Meets ISO 80601-2-55
• Designed and manufactured according to EN ISO 13485
• Accurate and reliable fast response
• Resistant to N2O
• Excellent signal stability
• High product quality
• Short delivery times
• Technical support
• Made in Germany