This Flat 13 Core Cable Shielded is intended for 13 Lead ECG Cable.13 leads ECG monitor cable can be done by a 11 lead Round cable or 13 conductor Flat cable.
Based on the same specification,there is Flat 10 lead ecg cable for option.
Conductor
Material
Tinned copper
Specification - 19/0.1TC
Insulation
Material HDPE
Nom.Thickness 0.25
Nom.O.D. ø0.95士0.05
Colour - White/Green/Black/Red/Brown/Blue/Purple/Gray/Orange/Yellow….
Conducting Layer
Material Semiconductive PVC
Nom.Thickness 0.1
Specification ø1.15士0.08
Colour Black
Shield
Material Tinned copper
Specification 43士3/0.08TC
Coverage >90%
Jacket
Material TPU 5040
Nom.Thickness 0.27
Nom.O.D. ø2.0士0.10*26.5士0.8
Colour White
Properties :
1. Hipot Testing: wire & wire, wire &shield: AC1000V , leakage current 0.25mA/MIN, and no breakdown, conductor and the outer Jacket :DC5000V, the leakage current is 0.25mA/S, and there is no breakdown.
2.Insulation Resistance: DC500V insulation above 100MΩ (test sample @3 meters).
3. Bending test:
The load is 200g, the lifting height is 300mm, the rounded corner of the fixture is R25.4MM, the swing speed is 30 times/min (2*90°),
More than 5000 times, no short circuit, no open circuit and damage to the insulation and Jacket.