The Safe AQ Smart Blood Glucose Monitoring System is designed for the quantitative measurement of glucose in fresh capillary whole blood samples taken from the fingertip and in venous whole blood samples. The Blood Glucose Monitoring System is for use outside the body only (in vitro diagnostic use) for self-testing and professional use as an aid in the management of diabetes.
Test principle
A blood glucose test is based on measurement of electrical current caused by the reaction of glucose with the reagents (special chemicals) on the electrode of the strip. The blood or control solution sample is drawn into the tip of the test strip through capillary action. Glucose in the sample reacts with the special chemicals and generates electrons, which produce electrical current. The blood glucose meter measures the electrical current and calculates the glucose result. The glucose result is displayed by your meter as mg/dL or mmol/L.
Features (Technical advantages)
1) FAD-GDH system : It is not easily affected by oxygen molecule in the blood.
2) User-friendly operation system: no coding & automatic test strip ejectionin the management of diabetes.
Specification
Blood volume - 0.6μL
Sample type - Capillary whole bloodyenous whole blood
Calibration - Plasma equivalent
Meter storage/transportation condition - -20℃~55℃
Dimension - 84*60*26(mm)
Power source - 3V DC, 50mA, 2 AAA alkaline batteries
Operating condition - 10℃~35℃; ≤80% RH
Measurement units - mg/dL or mmol/L
Measurement range - 20~600 mg/dL or 1.1~33.3mmo/L
Shelf life - 10 years(estimated by test 7 times per day).
During use, user should maintain the product
refer to this user manual’s requirements.