Advanced safety design able to stop infusion and alarm even if the CPU damage,the medical staff can ues it safely.
Precision Better than ±2%;
(but 5ml syringe and <1ml/h: Better than ±8%)
Parameter range 1, Flow-rate: 0.1~1200.0 ml/h, 0.1ml/h step;
Dose: 0.1~999.9 mg, 0.1mg step;
mg/kg/h: 0.001~999.999 mg/kg/h, 0.001 mg/kg/h step;
Body-weight: 1~300 kg, 1kg step;
Bolus volume: 0.1~99.9 ml/h, 0.1ml step
Alarm - 1, Occlusion,0~175KPa, 8 grades of sensitivity;
2, Pressure sensor is abnormal;
3, Push rod fall off;
4, The drug running out;
5, The injection will finish;
6, The motor rotation fault;
7, Overtime pause;
8, Low battery;
9, Battery failure;
10, Battery temperature is too high;
11, Mains missing tips;
12, MCU fault, and stop injection.
Data function 1, Memory of parameter: Records the latest injection parameter for next operation;
2, Historical records: records latest 1000 injection events.
(what/when/why,flow-rate, etc.)
Safety classification - Class I; Internal power supply, CF type, IPX4 equipment
Power supply - AC 100~240V, 50/60Hz, Built-in rechargeable Li-battery
Power consumption:20VA
Working time of battery: 10h@5ml/h