Three-step defibrillation process
Two-button operation
Extensive voice and visual prompts for the operator
Lock-out protection to prevent inadvertent defibrillation
Charge Time 8 sec. to 150J
10 sec to 200 J (New battery)
Analysis Time 9 sec.
The maximum time from the initiation of rhythm analysis to readiness for discharge with a new battery. Less than 30 seconds
The maximum time from the initiation of rhythm analy sis to readiness for discharge after 6 shocks. Less than 35 seconds
The maximum time from initially switching power on to readiness for discharge. Less than 25 seconds
Audible Prompts 20 audible prompts
Visual Prompts LED prompts
Controls Two buttons- On/Off, Shock
Output Energy Accuracy +15% into any impedance from 25 to 175Q
Output disabled when the patient impedance is outside Limits 20Q to 200Q
Shelf Life (25C±15C)
10 years (5 years storage + 5 years standby)5 years standby (after installation)
Physical
Operating Temperature 0Cto 40C
Operating Humidity
Relative humidity between 30% and 95%(non-condensing)
Storage Temperature
(without battery) -20C to 55C
Storage Humidity
(without battery) Up to 93% (non-condensing)