The only PMA approved excimer laser system and disposable catheters for the treatment of peripheral and coronary arterial disease. Photoablation is the use of light to break down, vaporize and remove matter. The CVX-300 excimer laser system facilitates photoablation via a variety of Philips laser catheters to remove binding arterial plaques.
Indicated treatments
Removing lesions comprising atheroma, fibrosis, calcium, thrombus and neointimal hyperplasia in the coronary and peripheral vasculature, and transvenous removal of problematic pacing and defibrillator leads.
Peripheral procedures
The CVX-300 can be used in the following peripheral procedures: treatment of infrainguinal stenoses and occlusions; critical limb ischemia (CLI); total occlusions crossable by guidewires
Coronary procedures
The CVX-300 can be used in the following coronary procedures: moderately calcified stenoses; occluded saphenous vein bypass grafts; long diffuse disease; restenosis in 316L stainless steel stents, prior to the administration of intravascular brachytherapyl; lesions that previously failed balloon angioplasty; balloon-refractory and balloon-resistant lesions; ostial lesions.
Specifications
System specifications
Active medium
XeCI
Weight
295 kg
Height
89 cm (unit), 18-23 cm (control pad)
Pulse width
125-200 ns, FWHM
Repetition rate
25-80 Hz
Length
125 cm
Width
62 cm
Catheter output fluence
30-80 mJ/mm2
Wavelength
308 nm