200 tests/hour (330 with ISE)
Washing station
Refrigerated compartment for reagents
Loading capacity for 40 samples
System Function:
Automatic, discrete, random access,
Bench-top, STAT sample priority
Throughput: Constant 200 tests/hour, up to
330 tests/hour with ISE
Measuring principles: Absorbance
photometry, Turbidimetry, Ion Selective
Electrode technology
Methodology: End-point, Fixed-time, Kinetic,
optional ISE single/dual reagent chemistries,
monochromatic / bi-chromatic
Open system with user defined profiles and
calculation
Reagent/Sample Handling:
Reagent/Sample tray: 40 positions for
reagents and 40 positions for samples in 24-
hour refrigerated compartment (2~12 ºC)
Reagent volume: R1: 10~350μl, step by 1μl;
R2: 10~200μl, step by 1μl
Sample volume: 2~45μl, step by 0.1μl.
Primary tubes and different sample cups.
Reagent/Sample probe: Liquid level detection,
vertical & horizontal collision protection and
inventory checking, reagent pre-warming
Probe cleaning: Automatic washing for
interior and exterior. Carry over < 0.1%
Automatic sample dilution: Pre-dilution and
post-dilution