Nurse call systems generally rely on LED Indicator and Tone Units to guide staff to the source of the call as quickly and efficiently as possible.
Units are installed at double doors and within corridors at strategic locations to provide 'follow-the-light' direction using the LED function, whilst the tones function acts as an audible guide and suppléments the sounder at nurse base indicators.
It may also be appropriate to install a LED Indicator and Tone Unit within staff rooms, ward kitchens and clean/dirty utility rooms.
Units continue to operate until the call is cancelled at the bedhead.
Facilities and Features
Facilities:
Illuminated LED
Tone Output
Features:
Intuitive and easy to use with similar appearance and system of operation to other Fusion -IP products
smartsync secure ‘two -way’ health -check for unrivalled system integrity. Faulty or missing devices reported within 2 minutes of failure Pre -programmed devices with ‘push -to-pair’
connectivity for quick set -up without the need for commissioning software
Full HTM 08:03 compliance
Steps to Adjust the Tone Unit Volume
Remove the screws located at the top and bottom of the LED Tone Unit.
Carefully unclip and remove the back cover.
Insert a terminal screwdriver in the Pot and rotate as follows:
Anti -clockwise - increase volume
Clockwise - decrease volume
Carefully clip the back cover into the correct position.
Insert the screws in the original location and tighten.