Freely confi gurable by user - without software installation
Clear alarm identifi cation
Automatic PDF report generation
Everything in one logger: confi guration fi le, PDF report, instruction manual and calibration certificate
High storage capacity: 8,000 data point memory
Manipulation - proof
Conforms to GxP, EN 12830 and FDA 21 CFR Part 11 / GAMP 5
Application
Monitoring of the cold chain during transport of sensitive freight such as pharmaceuticals, foods and technical products. The logger is designed for single use only.
Technical Specifications
Type of logger - Single use
Temperature sensor - NTC thermistor
Accuracy - ±0.5 °C (-30...70 °C)
Resolution - 0.1 °C
Range of application (humidity) - < 80 (3 / 0 R H
Protection - IP65 (when placed in plastic bag)
Weight - Approx. 10 g
Battery - CR2032 (not replaceable)
Dimensions without bag - 80(L) x 43(W) x 2.5(D) mm
Dimensions with bag - 105(L) x 55(W) x 2.5(D) mm
Logging interval - 0.5/5/10/30/60/90/120 min
Start delay - 0.5/5/10/30/60/90/120 min
Alarm type - Deactivated, single or cumulative
Storage period - 12 months
Storage capacity - 8192 data points
MARK function - Marking of up to 8 points during transport
Operation - START/STOP button, MARK button
LED indicators - LED «Low» flashes red
when value below alarm point
LED «Hi» flashes red
when value above alarm point
LED <, OK» flashes green during data logging
LED ((Battery» flashes red when low battery