The highest performance OEM module, Advantage MX, has unsurpassed motion artifact rejection and has been tested to extreme levels of simulated patient tremor/ shivering and travel on highway and gravel roads. Advantage MX acquires measurements quickly and reliably, even in difficult situations such as patient or vehicle motion.
Ultra-high Performance for the most challenging clinical applications
Advantage RMT-Max (Maximum Performance R-wave Motion Tolerance)
3rd Edition and RoHS Compliant
TTL communication protocol
Best motion tolerance for use in extreme transport environments
Physical Dimensions
Width 2.5 in [63.5 mm]
Length 3.1 in [78.7 mm]
Height 0.87 in [22.1 mm]
Weight 0.22 lbs [100 g]
Measurement
Technique Oscillometric
NIBP Accuracy Meets ANSI/AAMI SP10, EN 1060-4, ISO 81060-2
Systolic Range
Adult: 30-255 mmHg
Neonatal: 30-135 mmHg
Diastolic Range
Adult: 15-220 mmHg
Neonatal: 15-110 mmHg
Pulse Rate Range
Adult: 30-240 BPM (Beats per min)
Neonatal: 40-240 BPM
Pulse Rate Accuracy ±2% or ±3 BPM, whichever is greater
Transducer Accuracy ± 3 mmHg over full range in operating conditions
Operating Modes Manual
Operating Conditions 0˚C to 55˚C, 15% to 90%, non-condensing humidity
Storage Conditions -20˚C to +65˚C, up to 90% RH non-condensing
Electrical
Operating Voltage 12 VDC ± 5% (High), 6 VDC ± 5%(Low)
Power Consumption Using 6 VDC Nominal
Sleep: 300 mW
Idle: 380 mW
Measurement: 1.5 W
Inflation: 4.5 W (High)
Communication Protocols TTL (5V)
Safety & Regulatory Standards IEC 60601-1, IEC/EN80601-2-30, EN60601-1-2:2007, ISO 81060-2:2009