Colorimetric process analyzer for the automatic, continuous measurement of phosphate.
Measuring range: 0.01 to 10 ppm (mg/L) PO4
Based on colorimetric measurement principle according to ISO 6878
Complete system including measurement and control electronics, photometer, flow indicator, reaction chamber, reagent dosing system and reagent containers.
No interferences with silica
Control of sample flow and reagent fill level.
Automatic cleaning module available as option.
Factory tested and calibrated. Provides immediate results.
Optional accessory:
Cleaning module
Communication interfaces (RS-232, Profibus, Modbus, 3rd Signal Output, USB)
Specifications
Phosphate (PO4) Measurement -
Measuring range - 0.01 to 10 ppm
Reproducibility - ± 0.01 ppm or ± 2.5%
whichever is greater
Measurement time - 7 minutes
Cycle time - 10 minutes
Power Supply -
Voltage - 100-240 VAC (±10%), 50/60Hz
or
10-36 VDC
Sample conditions -
Flow rate - approx. 10 l/h
Temperature - up to 50° C
Inlet pressure - 0.15 to 2 bar
Outlet pressure - pressure free
Dimensions -
Panel - 400 x 850 x 150 mm