Monitor for continuous measurement of pH or redox (ORP) in potable water and effluents.
Complete system mounted on PVC panel:
Transmitter AMI pH-Redox in a rugged aluminum enclosure (IP 66).
Flow cell M-Flow 10-3PG with temperature sensor (Pt1000).
Factory tested and calibrated. Ready for installation and operation.
For use with combined sensors with reference elctrodes for various sample conditions:
Swansensor pH Standard or ORP Standard
Swansensor pH SI or ORP SI
Swansensor pH AY or ORP AY
Swansensor pH FL or ORP FL
Optional accessory:
Spray nozzle for sensor cleaning
Swansensor deltaT flow meter
Swan AMI Relaybox
Specifications
pH Measurement - with Swansensor ST/AY
Measuring range - pH 1 to 13
Resolution - 0.01 pH
Reference temperature - 25° C
Automatic temperature compensation
according to Nernst. -
ORP (Redox) measurement - with Swansensor ST/AY
Measuring range - -400 to 1'200 mV
Resolution - 1 mV
Temperature measurement -
Measuring range - -30 to 130°C
Resolution - 0.1° C
Power Supply -
Voltage - 100-240 VAC (±10%), 50/60Hz
or
10-36 VDC
Sample conditions -
Flow rate - 4 to 15 l/h
Temperature - up to 50° C
Inlet pressure - up to 1 bar
Outlet pressure - pressure free
Dimensions -
Panel - 280 x 850 x 150 mm