High precision monitor with three electrode Oxytrace G sensor for the continuous measurement of trace levels of dissolved oxygen in high purity water.
Integrated auto verification with faraday electrode setup.
Mounted on a small panel (width: 400mm, height 420mm)
Range: 0-20 ppm or 0-200 % saturation (at 25°C).
Automatic air pressure and temperature compensation.
Short rinse-down time due to integrated guard.
Faraday electrode and control unit for the automatic or manual verification.
Factory tested, ready for use.
Specifications
Complete system mounted on stainless steel mounting panel.
Transmitter in aluminum housing (IP 66) with 2 signal outputs for measuring value or for continuous control, 2 relays as limit switch or controller or timer, 1 alarm relay as summary alarm contact, 1 input for potential-free contact.
Flow cell QV-Flow PMMA OTG QED made of acrylic glass with needle valve, faraday electrode and digital sample flow meter on mounting angle made of stainless steel.
Faraday calibrator for the automatic verification of the instrument.
Measuring range with Swansensor OXYTRACE G (at 25 °C):
- 0 - 20.0 ppm O2 or
- 0 - 200% saturation
Process connections:
- Inlet: Swagelok ¼" tube adapter
- Outlet: G ½” adapter for flexible tube Ø 20 x 15 mm, pressure-free
Power Supply
-100-240 VAC (±10%)
-10-36 VDC
For further information see data sheet.