Air mode and baby mode controlled by micro-computer;
Set temperature, air temperature, baby temperature are displayed separately;
Natural air flow humidity;
Inclination of bassinet is adjustable;
Failure alarm indication;
Weighing system and Monitor shel fare optional.
PowerSupply: AC220V-230V/50Hz or AC110-120V/50-60Hz
Mode of control: Air mode and baby mode controlled by micro-computer
Control range of air temperature: 25 °C-37 °C(Override mode:37-39 °C)
Control range of baby temperature: 34 °C-37 °C (Override mode:37-38 °C)
Accuracy of skin temperature sensor: ±0.3 °C
Variability of temperature ≤0.5°C
Uniformity of temperature≤0.8°C
Inclination of bassinet±12 °
Noise inside hood: ≤55dB (A) (Environment noise≤ 45 dB(A))
Failure Alarm:
OverTemperature Alarm
Deviation Alarm
Sensor Failure Alarm
Fan Motor Alarm
PowerFailure Alarm
System Alarm