Teco Diagnostics is proud to introduce our complete line of Urinalysis products. From our world renowned Urine Reagent Strips (URS) to our semi-automated Urine Strip Readers.
Method:
Color Indicator
Testing Procedure:
Manual
Storage Temperature:
15‐30°C
Expected Values/Sensitivity:
Glucose: Concentrations as little 0.1g/dL may be significantly abnormal if found
consistently (100mg/dL sensitivity)
pH: Newborn: 5.0‐7.0; Thereafter: 4.5‐8.0 ; Average: 6.0 (quantitative differentiation of
values to one unit)
Protein: 1‐14mg/dL may be excreted by the normal kidney (15mg/dL sensitivity)
Limitations of Procedure:
Glucose: Moderate amounts of ketone bodies (40mg/dL or greater) may decrease color
development in urine containing small amounts of glucose (75‐125mg/dL). However,
such concentration of ketone simultaneously with such glucose concentration is
metabolically improbable in screening. The reactivity of the glucose test decreases as the
specific gravity and/or ascorbic acid of the urine increases. Reactivity may also vary with
temperature.
pH: If proper procedure is not followed and excess urine remains on the strip, a
phenomenon known as “running over” may occur, in which the acid buffer from the
protein reagent are runs onto the pH area, causing a false lowering in the pH result.
Protein: False positive results may be obtained with highly alkaline urine. Contamination
of the urine specimen with quarternary ammonium compounds may also produce false
positive results.
Expected Shelf Life:
90 days at 15‐30°C once opened
18 months at 15‐30°C unopened