For the direct, colorimetric determination of calcium in human serum or urine on TC Matrix analyzers.
Number of Tests:
580 tests
Testing Procedure:
Automated
Wavelength:
578 nm
Linearity:
20 mg/dL
Expected Values:
8.5-10.5 mg/dL
Children under 12 usually have high normal values which decrease with aging.
It is strongly suggested that each laboratory establish its own range of expected values.
Reagent Deterioration:
The reagent should be discarded if: (1) Turbidity has occurred; (2)The reagent fails to
meet linearity claims or fails to recover control values in the stated range.
Limitations of Procedure:
Substances that contain calcium or complex calcium should not come in contact with the
test specimen. Examples: EDTA, citrate, oxalate, and fluoride.
Specimens from patients receiving bromsulfophthalein (BSP) or EDTA should not be
used.
Lipemic or hemolyzed samples require a serum blank.
Contamination of glassware with calcium (usually from detergents) will adversely affect
the test. Use acid-washed glassware or plastic tubes.