The ST-IS is a cGMP Class A/ISO 5 modular isolator system specifically designed for Sterility Testing, which guarantees the safest working conditions during Quality Control Operations. A compact configuration (CST-IS) is available.
OPERATION
ST-IS is an aseptic isolation system designed to provide the best environmental and working conditions for conducting sterility tests according to Pharmacopeia.
ST-IS consists of two chambers with different characteristics and functionalities. The first step of the workflow involves placing the material for sterility testing into the Transfer Chamber, followed by the execution of the decontamination process using the integrated VPHP system. Once the process is complete, the material can be transferred to the Processing Chamber without compromising the aseptic level of the Processing Chamber, which has been previously subjected to the decontamination process.
The Working Chamber is equipped with a state-of-the-art integrated peristaltic pump, allowing for simple and reliable execution of the test.
MAIN FEATURES
GUARANTEED ASEPSIS
The ST-IS assures the highest air quality in the isolated environment.
AUTOMATIC LEAK TEST
The ST-IS is equipped with a routine automatic leak testing system which performs an independent leak test cycle for each chamber according to the pressure change method (5.2) ISO 10648-2(E).