BPH-Photon. 7 sec. Designed for the incubation and automatic readout of Biological Indicator for Steam sterilization processes. Final readout: 7″.
FDA granted
Product description
Terragene® Bionova® Photon Auto-reader Incubator (BPH) has been designed for the incubation and automatic readout of the Bionova® Photon line of Self-contained Biological Indicators (SCBIs) appropriate for monitoring Steam Sterilization Cycles. Bionova® Photon Auto-reader presents two independent readout positions. Each position automatically detects whenever a SCBI is placed for incubation and delivers a readout result after a few seconds using advanced fluorescence techniques.
Bionova® Photon Auto-reader allows easy and rapid detection of positive and negative SCBIs. A positive result can also be evidenced by culture medium color change when performing extended incubations. Read the instructions for use of each SCBI for more information. The option to perform or not an extended incubation depends on the internal protocols of
each laboratory or hospital.
Bionova® Photon Auto-reader has built-in USB, Wi-Fi, and Bluetooth® technology connectivity capabilities. The electronic tickets (e-tickets) created for each readout process can be accessed using a compatible device thus allowing easy documentation compliance and traceability.
Instructions for use
1) Place the Bionova® Photon Auto-reader on a firm surface, free from vibrations, away from direct sunlight, currents of hot or cold air, chemical and corrosive or flammable substances.
Do not place the Incubator in a way that disconnection of the power supply’s AC plug from mains could be difficult.