The ground-breaking VitaloLAB™ combines power and precision in one unique full pulmonary function testing system. Designed to be compliant with latest international guidelines for accuracy, repeatability and grading. The impressive VitaloLAB PFT system gives you full spirometry, provocation, DLco (RGA), lung volume measurements, lung clearance index (LCI) and respiratory muscle assessments as standard plus many more features that will blow you away.
The ideal PFT lab solution with complete testing options available
Powered by the first-in-class ComPAS2 software.
Stay connected with simple workflows with ComPAS2 Task Manager
Essential Performance Test Limits
Data will be transferred by USB from the device PCBs to the on-board computer. The output from the DLco Gas Analyser must stay within the predetermined limits (see below) both in normal and single fault conditions.
Carbon Monoxide (CO) values in DLco Gas Analyser: +/- 30 counts
Methane (CH4) values in DLco Gas Analyser: +/- 75 counts
Flow Detection Principal Heated Lilly type pneumotachography
Back pressure Less than 0.15kPa/L/second @ 14L/s, complies with ISO26782:2009)
Volume detection Flow integration sampling @ 100Hz
Volume Accuracy Within ±2.5% or ±75 mL
Voltage/Frequency 80-240V; approximately 50/60 Hz
Flow accuracy Within ±2% over the range of -14 to +14 L/s