The ground-breaking VitaloQUB™ combines power and precision in one unique full pulmonary function testing system. Designed to be compliant with latest international guidelines for accuracy, repeatability and grading. The impressive VitaloQUB PFT system gives you full spirometry, provocation, DLco (RGA), lung volume measurements, lung clearance index (LCI) and respiratory muscle assessments as standard plus many more features that will blow you away.
Expand your testing capabilities by adding whole body plethysmography and airways resistance measurements.
Powered by the first-in-class ComPAS2 software.
Bench seat to support up to 250kg and low step height for patient comfort.
Dedicated patient incentive screen to help patient technique for optimal results.
The VitaloQUB is used in conjunction with either the VitaloLAB and VitaloROV+.
Essential Performance Test Limits
Data will be transferred by USB from the device PCBs to the on-board computer. The output from the DLco Gas Analyser must stay within the predetermined limits (see below) both in normal and single fault conditions.
Carbon Monoxide (CO) values in DLco Gas Analyser: +/- 30 counts
Methane (CH4) values in DLco Gas Analyser: +/- 75 counts
Flow Detection Principal Heated Lilly type pneumotachography
Back pressureLe ss than 0.15kPa/L/second @ 14L/s, complies with ISO26782:2009)
Volume detection Flow integration sampling @ 100Hz
Volume Accuracy Within ±2.5% or ±75 mL
Voltage/Frequency 80-240V; approximately 50/60 Hz
Flow accuracy Within ±2% over the range of -14 to +14 L/s