Incubators and radiant warmers are used to maintain the body temperature of newborn infants. This is best done so that the energy expended for metabolic heat production is minimized. The heat output of these devices is usually regulated by servocontrol to keep the skin temperature constant at a site on the abdomen where a thermistor probe is attached.
Product Description
■Brief Introduction:
Air mode control, set temperature and air temperature are displayed separately;
Natural air flow humidity
Inclination of bassinet is adjustable
Failure alarm indication
RS-232 connector
Oxygen input interface
■Standard configuartion:
HB-YP90:the hood ,the base ,the controller ,I.V pole ,shelf.
■Optional parts:
> 37°C temperature override mode function, vertical height adjustment stand(VHA stand),
Weighing system, Neonate bilirubin phototherapy equipment model: HBXHZ-90L (fixed model), Baby mode and SPO2.
■Specification:
Power requirement: AC220V-230V/50Hz or AC110-120V/50-60Hz, 850VA
Mode of control: Air mode control
Air temperature control range: 25°C -37°C (override mode: 37-39°C is optional)
The temperature difference between incubator temperature and average incubator
temperature(under steady temperature condition)≤0.5°C
Temperature uniformity: 0.8°C (mattress is on the horizontal position)
Inclination of the bassinet: ±12°
Noise inside hood: 45dB(A)
Failure alarm: over temperature alarm, deviation alarm, sensor failure alarm, fan motor alarm, power failure alarm, system alarm and so on.
The average value of total irradiance for bilirubin Ebi max of phototherapy equipment:
≥ 640pW/cm2 ( XHZ-90 ) • ≥1300pW//cm2 ( HBXHZ-90L )