Defibrillators have long used a monophasic waveform, where current flows in one direction from one electrode to the other, stopping the heart momentarily, and allowing the basic sinus rhythm to be restored.
Product Description
Physical characteristics:
Main unit: Portable
Technical specifications:
(1)Defibrillator type Manual, Synchronized, Asynchronized
(2)Energy accuracy +1%
(3)Energy selection: 0, 3, 5, 7, 10, 20, 30, 50, 100, 200, 300, 360 joules (nominal at a resistance of 50)
(4)Charging time: Maximum 10 seconds to 360 Joules
(5)Standard Adult/Pediatric Paddles: Reusable external adult paddles (pediatric paddles integrated)
Power requirements
(a)AC Power supply 100V~240V,50Hz/60 Hz
(b)Vehicle Voltage(Option) DC 12V
Internal Battery Backup
(1)Battery type: Rechargeable lead-acid battery (12V 2.0Ah X2)
(2)Charging time: Time-Minimum of 4 hours for full charging
(3)Battery charging: During AC Power Supply operation of the device
(4)Charging /Low Battery Indication: Red LED
(5)Battery Capacity: 100% of the battery will provide 120-min monitoring;
Display
-LCD Dimensions 7-inch
-Type High resolution true-color: graphic
-Resolution 480X234
ECG
ECG lead: AVR, AVL, AVF, V1~V6
Protection: With stand 4000vAc/50Hz voltage in isolation and work against electrosurgical
knife and defibrillation
Respiration
Measurement method: Thoracic impedance
NIBP
Working mode: Manual/Automatic/stat(5 minutes)
Method: Automatic oscillation
Cuff type: standard Adult; optional Pediatric/Neonatal
Temperature
Measurement range 20~45℃
Resolution 0.1
Pulse Oximetry(spO2)
Display :spo2 value, pulse histogram, waveform, Pulse