Power requirements of the machine room
Voltage conditions: three phase 380V±10%;
Power frequency: 50Hz±1Hz;
Power resistance: ≤0.25Ω;
The power supply capacity: ≥50kVA.
High frequency and high pressure generator
High frequency and high voltage generator: maximum power 50kW, maximum voltage 150kV;
Invert frequency: 60kHz;
Photography kV: 40 to 150kV, step 1 kV;
Photography mA: 20 ~ 630mA, 16 gear adjustable;
Exposure time s: 0.002 to 10s, 34 gear adjustable;
Auto exposure: Yes;
Fault self diagnosis function: Yes;
Anti overload function: Yes.
X ray tube assembly
Maximum working voltage: 150kV
Maximum anode thermal capacity: 150kHU
Focus nominal value: small focal 0.6cm, big coke 1.2cm
Maximum instantaneous power: small focal 20kW, large coke 50kW
Anode speed: 2800r/min
Multifunction U arm frame
The rotation angle of the U arm: -30° to +120°;
Vertical movement of U arm: 60cm to 1700cm SID: 100cm ~ 180cm
Detector rotation: -45°to +45°
X ray tube ball rotation: -180° to +180°, 90°section, mechanical limit
Flat panel detector
Type of detector: amorphous silicon flat panel detector
Image size: 17 x 17
Pixel matrix: 3027 * 3027
Pixel size: 139 mu m
Spatial resolution: 3.6lp/mm
A/D Transformation: 14-bit
Image acquisition time: less than 3s
Image processing time: 5~8 s
Image acquisition workstation
CPU frequency: Intel I5
Host memory: 8G high speed memory
Hard disk: large capacity and high speed hard disk 500G
Display: 24 inch liquid crystal display
Network interface: 1000M
Interface protocol: DICOM3.0
Packing Size
2350*1000*950MM
1050*1050*750MM
G.W 630KG
N.W 540KG